Hardness – LaMotte Fresh Water Aquaculture User Manual
Page 22
22
Hardness
DESCRIPTION
CODE
*Hardness Reagent #5
*4483-E
Hardness Reagent #6 Solution
4485-E
Hardness Reagent #7
4487DR-H
Test Tube, 5-10-12.9-15-20-25, glass, w/cap
0608
Direct Reading Titrator, 0-200 Range
0382
Pipet, 0.5 mL, plastic
0353
*WARNING: Reagents marked with an * are considered to be potential health hazards.
To view or print a Material Safety Data Sheet (MSDS) for these reagents go to
www.lamotte.com. To obtain a printed copy, contact LaMotte by e-mail, phone or fax.
The Titrator is calibrated in terms of Total Hardness expressed as parts per million (ppm)
Calcium Carbonate (CaCO
3
). Each minor division on the Titrator scale equals
4 ppm CaCO
3
.
ANALYSIS OF HARDNESS IN SALT WATER
1
3
2
Use the 0.5 mL
pipet (0353) to
add 0.5 mL of the
salt water to the
test tube (0608).
Fill the test tube
to the 12.9 mL
line with distilled
water (a 1 to
25.8 dilution).
Follow Steps 2 through
9 under Hardness Test
procedure. Multiply
reading by 25.8. Record
as ppm Total Hardness
CaCO
3
.
When sea and estuarine waters containing very high levels of mineral salts are to be tested,
the sample must be diluted before titration. This test set is supplied with a calibrated pipet
for perforning the simple, convenient dilution described below: