beautypg.com

LaMotte PERACETIC ACID TEST KIT User Manual

Peracetic acid test kit, Use the instruction in indesign

background image

PERACETIC ACID TEST

KIT

CODE 7191-01

QUANTITY

CONTENTS

CODE

30 mL

*Sulfuric Acid, 1:1

*6141WT-G

15 mL

Ferroin Indicator

6410-E

2 x 30 mL

*Hydrogen Peroxide Titrant

*5650LWT-G

30 mL

*Potassium Iodide 20% Solution

*6521-G

60 mL

Peracetic Acid Titrant

S-6155-H

1

Test Tube, 5-10-25 mL, plastic, w/cap 0715

1

Pipet, 0.5 mL, plastic

0353

*WARNING: Reagents marked with an * are considered to be potential health
hazards. To view or print a Material Safety Data Sheet (MSDS) for these
reagents go to www.lamotte.com. To obtain a printed copy, contact LaMotte by
e-mail, phone or fax.
To order individual reagents or test kit components, use the specified code
number.
NOTE:

Do not use for tapwater or less than 20 ppm peracetic acid.

PRO CE DURE

HYDROGEN PEROXIDE

1.

Fill test tube (0715) to 10 mL line with sample water.

2.

Add 10 drops of *Sulfuric Acid, 1:1 (6141WT). Swirl to mix. Make sure
solution is thoroughly mixed.

3.

Add 1 drop of Ferroin Indicator (6410). Swirl to mix. Make sure solution is
thoroughly mixed. Solution will turn pale orange-pink if hydrogen
peroxide is present.

4.

While gently swirling tube, add *Hydrogen Peroxide Titrant (5650LWT)
one drop at a time until pale orange-pink color changes to colorless or pale
yellow. Make sure solution is thoroughly mixed. Count the number of drops
added. Hold bottle vertically.

5.

Discard titrated sample and rinse tube out with sample water.

6.

Multiply the number of drops used in Step 4 by 50 (fifty). Record as ppm
Hydrogen Peroxide.

Use the instruction

in