LaMotte Water Analysis Kit for Home Brewers User Manual
Page 5
1
2
3
5
6
4
6
CALCIUM HARDNESS
Fill tube (
4488) to line
with water sample.
Add
6 drops of
*Sodium Hydroxide
Reagent (4259)
and
swirl to mix.
Add
1 Calcium Hardness
Indicator Tablet (T-5250)
and
swirl to dissolve tablet.
Sample will turn
Red
if
Hardness is present.
Proceed to next step.
If sample is
Blue
, Calcium
Hardness is 0 ppm.
Record as ppm
CALCIUM
Hardness (CaCO
3
ppm).
To convert to calcium
(Ca
+2
) multiply the test
result by 0.4.
MAGNESIUM HARDNESS
TOTAL Hardness value____minus CALCIUM Hardness value ____=
MAGNESIUM Hardness (CaCO
3
ppm)
To convert to magnesium (Mg
+2
) multiply the test result by 0.24.
Line
Equivalence
(ppm per drop)
Upper
1 drop = 10 ppm
Lower
1 drop = 20 ppm
Hold the
Hardness Reagent #7 (4487WT)
vertically to add one drop at a time
(
swirling and counting after each drop)
until color changes completely from
Red
to
Blue
and remains for 30 seconds.
Brush and rinse tube promptly after testing.
Multiply number of drops:
Upper line
multiply x 10
Lower line
multiply x 20
If the calcium hardness is greater than 500 ppm,
use the 0.5 mL pipet (0353) to transfer 1.0 mL
(two measures) of the water sample to the
test tube (4488). Fill tube to 10 ppm line with
distilled water. Follow Steps 2 through
5. Multiply the number of drops by 86.