Hardness tests (as caco – LaMotte Water Analysis Kit for Home Brewers User Manual
Page 4
HARDNESS TESTS (as CaCO
3
)
QUANTITY
CONTENTS
CODE
15 mL
*Sodium Hydroxide Reagent w/Metal Inhibitors
*4259-E
50
Calcium Hardness Indicator Tablets
T-5250-H
15 mL
*Hardness Reagent #5
*4483-E
100
Hardness Reagent #6 Tablets
4484-J
60 mL
Hardness Reagent #7
4487WT-H
1
Test Tube, w/cap
4488
1
Pipet, 0.5 mL, plastic
0353
1
2
3
5
6
4
6
TOTAL HARDNESS
Fill tube (
4488) to the line with
sample water.
Add
5 drops of
*Hardness Reagent
#5 (4483) and swirl
to mix.
Add
1 Hardness
Reagent #6 Tablet
(4484) and
swirl to
dissolve tablet.
Sample will turn
Red
if Hardness
is present. Proceed to next step.
If sample is
Blue
, Total Hardness
is 0 ppm.
Hold the
Hardness Reagent #7 (4487WT)
vertically to add one drop at a time
(
swirling and counting after each drop)
until color changes completely from
Red
to
Blue
and remains for 30 seconds.
Brush and rinse tube promptly after testing.
Line
Equivalence
(ppm per drop)
Upper
1 drop = 10 ppm
Lower
1 drop = 20 ppm
Multiply number of drops:
Upper line
multiply x 10
Lower line
multiply x 20
Record as
TOTAL
Hardness (CaCO
3
ppm)
If the total hardness is greater than 500 ppm,
use the 0.5 mL pipet (0353) to transfer
0.5 mL of the water sample to the test tube
(4488). Fill tube to 10 ppm line with distilled
water. Follow Steps 2 through 5. Multiply the
number of drops by 172.