HARDNESS TESTS (as CaCO

3

)

QUANTITY

CONTENTS

CODE

15 mL

*Sodium Hydroxide Reagent w/Metal Inhibitors

*4259-E

50

Calcium Hardness Indicator Tablets

T-5250-H

15 mL

*Hardness Reagent #5

*4483-E

100

Hardness Reagent #6 Tablets

4484-J

60 mL

Hardness Reagent #7

4487WT-H

1

Test Tube, w/cap

4488

1

Pipet, 0.5 mL, plastic

0353

1

2

3

5

6

4

6

TOTAL HARDNESS

Fill tube (

4488) to the line with

sample water.

Add

5 drops of

*Hardness Reagent

#5 (4483) and swirl

to mix.

Add

1 Hardness

Reagent #6 Tablet

(4484) and

swirl to

dissolve tablet.

Sample will turn

Red

if Hardness

is present. Proceed to next step.
If sample is

Blue

, Total Hardness

is 0 ppm.

Hold the

Hardness Reagent #7 (4487WT)

vertically to add one drop at a time

(

swirling and counting after each drop)

until color changes completely from

Red

to

Blue

and remains for 30 seconds.

Brush and rinse tube promptly after testing.

Line

Equivalence

(ppm per drop)

Upper

1 drop = 10 ppm

Lower

1 drop = 20 ppm

Multiply number of drops:

Upper line

multiply x 10

Lower line

multiply x 20

Record as

TOTAL

Hardness (CaCO

3

ppm)

If the total hardness is greater than 500 ppm,

use the 0.5 mL pipet (0353) to transfer

0.5 mL of the water sample to the test tube

(4488). Fill tube to 10 ppm line with distilled

water. Follow Steps 2 through 5. Multiply the

number of drops by 172.