TOTAL RESIDUAL CHLORINE AND COMBINED CHLORINE
1.

Rinse test tube (0106) with sample water. Fill to 10 mL line with
sample water.

2.

Add one *Chlorine DPD #1R Tablet (6999A). Cap and mix until
tablet disintegrates.

3.

Immediately insert test tube into Chlorine DPD Comparator (6978).
Match sample color to a color standard. Record as ppm Free
Available Chlorine.

4.

Add one *Chlorine DPD #3R Tablet (6905A). Cap and mix until
tablet disintegrates.

5.

Insert test tube into Chlorine DPD Comparator (6978). Match
sample color to a color standard. Record as ppm Total Residual
Chlorine.

Combined Chlorine, ppm =

Total Residual Chlorine - Free Available Chlorine

TOTAL RESIDUAL CHLORINE
*Chlorine DPD #4 Tablet determines Total Residual Chlorine and is used
where a differentiation in the type of chlorine is not required.
1.

Rinse test tube (0106) with sample water. Fill to 10 mL line with
sample water.

2.

Add one *Chlorine DPD #4 Instrument Grade Tablet (6906A). Cap
and mix until tablet disintegrates.

3.

Insert test tube into Chlorine DPD Comparator (6978). Match
sample color to a color standard. Record as ppm Total Residual
Chlorine.