beautypg.com

Air sampling apparatus, Air sampling pump model bd • code 1949, Instructions – LaMotte Air Pollution Sampling & Measurement User Manual

Page 7: Diagram 1

background image

AIR SAMPLING APPARATUS

AIR SAMPLING PUMP MODEL BD • CODE 1949

The Model BD Air Sampling Pump is designed especially for use with the
LaMotte Air Pollution Test Equipment. Among the many features of the
air sampling pump are an adjustable flowmeter, intake and exhaust
connections, impinger holder, on-off switch, four “C” cell batteries and
AC jack. For proper maintenance and to guarantee the life and
usefulness of the air sampling pump, the following operating procedures
are recommended.

INSTRUCTIONS

1.

Check battery capacity by turning switch to the “ON” position. If the
red LED on the front of the meter fails to come on, the batteries are
too weak to operate the pump. Adjust flowrate to sample at 2.0
Lpm. If indicator float of flowmeter vibrates up and down with a
deflection of more than 0.2 Lpm, or if air sampling pump cannot be
adjusted to sample at 2.0 Lpm, replace batteries (see Battery
Replacement). Alternatively the pump may be operated with an AC
adapter (1744).

Air Pollution Sampling & Measurement

7

AIR

SAMPLING

PUMP

IN

OUT

2.0

1.5

1.0

.5

LPM AIR

ON • OFF

BATTERY

OPERATION

7 ml

20 ml

30 ml

40 ml

5 ml

10 ml

15 ml

25 ml

CODE 0922l

AC Adapter

Jack

Impinger

Tube

(0922)

Outlet

Tube

Stopper

Assembly

Intake

Tube

Flexible Tubing

Pump Intake

Fitting

Flowmeter

Diagram 1