beautypg.com

LaMotte WaterLink Spin User Manual

Page 8

background image

8

GENERAL OPERATING PROCEDURES

METER

When a fi lled disk is placed in the chamber and the lid is closed, the meter

spins at high speed to distribute the sample to all 11 wells and to ensure that

the water empties from the center chamber. Next the meter slows to maximize

the pumping action of the stainless steel mixing beads as the reagents mix

with the sample water. Each reaction is then read at the proper time and

wavelength for that reagent system.

The blue indicator light located below the thumb hole on the meter housing will

indicate the status of the instrument. A steady blue light indicates that the

meter is connected to the power source and the computer and is ready to run
a test. A blinking blue light indicates that a test is in progress and that the

disk is spinning and the meter is taking readings. Do not open the lid while the

disk is spinning. The blue light will remain on when the meter is plugged into

the electrical outlet.

Care should be taken when closing the lid. Do not slam the lid. Wiring

between the lid and the body of the photometer passes through the hinge. The

meter will not run with the lid open.

SYRINGE

A plastic 3 mL syringe (Code 1189) is used to fi ll the disks. A precision tip

on the syringe fi ts into the fi ll hole on the disk. The syringe tip should not

be removed from the syringe. Syringes should be cleaned between water

samples. Pump air in and out of the syringe a few times to clear the previous

sample or rinse the syringe with a small amount of the next water sample

before fi lling it for testing.

DISKS

The WaterLink Spin uses a disk reagent system. The dried reagents are

packaged in single test amounts in a sealed, polystyrene disk. Stainless steel

mixing beads in the reaction chambers mix the sample water and the dried

reagents.

Chlorine/Bromine and Biguanide Disks
There are two disk series – Chlorine/Bromine and Biguanide. Each disk

includes a set of reagents for the series. Tests for all factors in the series are

performed at one time. It is not possible to isolate the well for a single factor

and perform a test for one test factor only.