beautypg.com

LaMotte WaterLink Spin User Manual

Page 14

background image

14

TROUBLESHOOTING

TROUBLESHOOTING GUIDE

Problem

Reason

Solution

Connection Error

Could not connect to

water analysis program.

Adjust program

settings. Be sure

program is running.

USB 2 port required.

Lid Open Error

Lid open.

Close lid and begin

again.

Test Series Cancelled

Error

Lid open, meter not

connected.

Close lid. Reestablish

connection.

Unexpected results

Did not use disk cover.

Retest. Use disk cover.

Unexpected low chlorine

results

High chlorine

concentrations will

bleach chlorine

reagents.

Check sample with

chlorine test papers

(Code 2987-G).

Disk is not spinning

Disk or cover pressed

too fi rmly onto hub.

Reseat disk and cover

on hub.

Results not as expected Meter out of calibration

Confi rm test results with

another method. Check

meter calibration with

Calibration Check Disk.

(not included)

HELPFUL HINTS

»

High Chlorine/Bromine - High sanitizer levels may cause the DPD

reagents in the disk to bleach out to a colorless or near colorless

solution. Wide Range pH/Total Chlorine Test Strips (2987-G) capable

of testing high concentrations of chlorine can be used to determine the

approximate level of sanitizer. Note: At high sanitizer levels, chloramines

could break through into the Free Chlorine test, the pH reaction may

turn purple and the Alkalinity reaction may turn yellow.

»

High Copper - Copper at levels above 1.0 ppm may cause the

Hardness test to read low.

»

Low Alkalinity - Algaecide treatments above recommended levels may

lower the alkalinity results.

»

Cyanuric Acid - Since Cyanuric Acid tests are temperature dependent,

the best results are obtained when sample temperatures are between

70 and 80 degrees. Below this range, results may read high, while in