LaMotte TRACER POCKETESTER TDS User Manual

Page 7

background image

7

BASIC OPERATION

Powering the TRACER
The Tracer uses four CR2032 Lithium Ion batteries. If the batteries are weak,
the

BAT

indicator will appear on the display. Press the ON/OFF key to turn

the TRACER on or off. The auto power off feature will shut the TRACER off
automatically after 10 minutes after the last button push.

Automatic Calibration
When the TRACER is turned on, it will enter the Automatic Calibration
mode.

SELF

and

CAL

will appear while the calibration is in progress. After the

calibration is completed, the

SELF

and

CAL

display icons will extinguish.

TDS Conversion Ratio
The TDS value is determined by multiplying the conductivity measurement
by a known conversion ratio factor. The meter allows the selection of a
conversion ratio factor in the range of 0.4 to 1.0. The selected ratio will vary
with application but is typically set between 0.5 and 0.7. In the salinity mode,
the ratio is fi xed at 0.5. The stored ratio factor will briefl y appear in the lower
temperature display when the meter is fi rst turned on or when changing the
measurement function to TDS.

To Change the TDS Conversion Ratio:
1. Turn the TRACER on.
2. Press and release the CAL/RECALL button twice.

The stored ratio will appear in the display.

3. Press the MODE/HOLD button to change the ratio

value in steps of 0.1.

4. When the desired ratio is displayed, press and release

the CAL/RECALL button to store the value and return
to the normal mode.

5. If no buttons are pressed for 5 seconds, the meter

will return to the measurement mode.