beautypg.com

LaMotte TRACER POCKETESTER TDS User Manual

Page 18

background image

18

TROUBLESHOOTING

Problem

Check

Action

Reading is frozen

HOLD mode

Press MODE/HOLD button
to exit HOLD mode

“BAT” message

Batteries low

Replace batteries

Meter will not
calibrate

Trapped air
bubbles

Tap probe or stir sample to
release air bubbles

Dirty probe

Clean probe

Damaged
probe

Replace probe

Contaminated
standards

Use fresh standards

Meter will not turn
on

Batteries low
or dead

Replace batteries

Battery
polarity

Replace batteries with correct
polarity

Unit will not respond
to any key presses

Internal fault

Perform hard reboot. Remove
batteries, hold ON/OFF
button down for 5 seconds,
replace batteries

“_oL_” message

Sample is out
of range

Dilute sample

This manual is related to the following products: