LaMotte TRACER POCKETESTER TDS User Manual
Page 18
18
TROUBLESHOOTING
Problem
Check
Action
Reading is frozen
HOLD mode
Press MODE/HOLD button
to exit HOLD mode
“BAT” message
Batteries low
Replace batteries
Meter will not
calibrate
Trapped air
bubbles
Tap probe or stir sample to
release air bubbles
Dirty probe
Clean probe
Damaged
probe
Replace probe
Contaminated
standards
Use fresh standards
Meter will not turn
on
Batteries low
or dead
Replace batteries
Battery
polarity
Replace batteries with correct
polarity
Unit will not respond
to any key presses
Internal fault
Perform hard reboot. Remove
batteries, hold ON/OFF
button down for 5 seconds,
replace batteries
“_oL_” message
Sample is out
of range
Dilute sample