CALIBRATION & ANALYSIS

CALIBRATION

Turbidity Standards

Only use AMCO or formazin standards with the 2020we/wi. StablCal

®

standards below 50 NTU should not be used to calibrate the 2020we/wi.

The diluent used in the StablCal

®

standards has a different refractive

index than traditional formazin standards and will affect the results.

The concentration of the calibration standard should be similar to the

expected concentration of sample that will be tested. The following

standards are available from LaMotte Company:
1480 0 NTU/FNU Standard, 60 mL (EPA or ISO)
1450 1 NTU Standard, 60 mL (EPA)
1453 1 FNU Standard, 60 mL (ISO)
1451 10 NTU Standard, 60 mL (EPA)
1454 10 FNU Standard, 60 mL (ISO)
1452 100 NTU Standard, 60 mL (EPA)
1455 100 FNU Standard, 60 mL (ISO)

Turbidity Calibration Procedure

The default units are NTU and FNU and the default calibration curve

is formazin as indicated by (F) in the Menu bar. A 2020we, which uses

NTU, will be used in the following examples. For the most accurate

results, a user calibration should be performed. The Japan Standard

calibration mode, as indicated by (J) in the Menu bar, should be used

only with Japanese Polystyrene Standards (0-100 NTU). To change the

settings see the Set Up Instructions on page 9.
For the most accurate results, perform a calibration over the smallest

range possible.

Use a calibration standard that, along with the blank,

brackets the range of the samples that will be tested. For example, if

the samples that are to be tested are expected to be below 1 NTU, more

accurate results will be obtained by calibration with a blank and a 1 NTU

standard as opposed to a blank and a 100 NTU standard.

The meter has fi ve measuring ranges:
0 – 11 NTU/FTU

11 – 110 NTU/FTU

110-300 NTU/FTU

300-600 NTU/FTU

600-4000 NTU/FTU

Each range can be calibrated with one point per range. (Six points

total - a blank plus one point in each of the fi ve ranges - if each range

is calibrated.) New calibration points will replace old calibration points

independently for each range. If one range is recalibrated, the meter will

36