beautypg.com

LaMotte 1200-NA COLORIMETER User Manual

Page 3

background image

NITRATE - NITROGEN

3

NOTE:

11.

9.

10.

NOTE:

At the end of waiting period,

an undissolved portion of

*Nitrate Reducing Reagent

may remain in bottom of the

tube without affecting results.

Use the 0.1 g spoon

(0699) to add two

measures of *Nitrate

Reducing Reagent

(V-6279). Cap.

Hold tube by

index finger and

thumb and mix by

12.

inverting approximately

50-60

times a minute for

4 minutes

. Wait 10

minutes

for maximum

color development.

10

MIN

UTE

S

Wipe tube dry.

Align the index line with the

arrow on the meter, insert

CAL

I200

COLORIMETER

• • • • • • • • • • • • • • • • • •

READ

tube into chamber.

Close the lid. Push

the

READ

button. Record

results as ppm

Nitrate Nitrogen.

To convert Nitrate Nitrogen

(NO -N) results to ppm Nitrate

3

(NO ), multiply by 4.4.

3

L Mott