beautypg.com

LaMotte 1200-NA COLORIMETER User Manual

Page 2

background image

NITRATE-NITROGEN TEST PROCEDURE -

CADMIUM REDUCTION METHOD

Read the 1200 Colorimeter Manual before proceeding. Carefully
wipe tubes dry before inserting into the colorimeter chamber.

NITRATE - NITROGEN

2

Insert the tube

into the

chamber, being

sure to align the

index line with

the arrow on

the meter.

Close the lid.

This tube is the

blank or zero.

8.

5.

3.

4.

Fill the Water

Sample Collecting

Bottle (0688) with

sample water. This

will be used to

dispense sample

water for the tests.

Rinse and fill a

colorimeter tube

(0290) to the 10

mL line with

sample water.

Cap and wipe

dry.

1.

2.

I200

COLORIMETER

• • • • • • • • • • • • • • • • • •

READ

bla

ZERO

7.

Remove tube from

colorimeter and

pour off 5 mL into

graduated cylinder

or similar. Discard

the remaining

sample.

6.

Pour the 5 mL

sample from a

graduated

cylinder or

similar into the

colorimeter

tube.

Push the READ
button to turn the

meter on. Press the

ZERO

button

and hold it for 2

seconds until

bla

is

displayed. Release

the button to take a

zero reading
(0.00 ppm).

Use the graduated

cylinder or similar to

measure 5 mL of

*Mixed Acid

Reagent (V-6278)

and add to tube.

5 mL

5

m

L

2

MIN

UTE

S

5 mL

Cap and mix. Wait approximately 2

minutes before proceeding to Step 9.

L Mott