SPECIFICATIONS
IN STRU MENT TYPE
Multi-wavelength filter colorimeter _ internal, non-removable filters
READ OUT
3
1
2
inch digit LCD; displays 0–100%T
READ ABLE RES O LU TION
±
1%T
READING STA BIL ITY
±
0.2%T within 5 seconds of turn-on to automatic turn-off
READING IN TER VAL
Approximately 30 seconds with automatic turn-off, resettable
MEA SURE MENT WAVE LENGTHS
1 (420nm), 2 (460nm), 3 (510nm), 4 (530nm), 5 (570nm), 6 (605nm); switch selectable
WAVE LENGTH AC CU RACY
±
1 nanometers
PHO TO MET RIC AC CU RACY
±
0.5%T
SAM PLE CHAM BER
Indexed; accepts 21 mm diameter flat-bottomed test tubes (capped)
SOURCE LAMP
Tungsten filament bulb, 10,000 hour life (est.), spare included, field replaceable
POWER RE QUIRE MENTS
Battery Operation: Field replaceable 1604 type (9V)
Line Operation: 120/220V, 50/60 Hz, 2VA, with optionally-available adapter
DI MEN SIONS
(W x D x H) 190 x 140 90 mm
7
1
2
x 5
1
2
x 3
1
2
inches
WEIGHT
2 lbs.
6
1985-01 • 10/01