GENERAL INFORMATION

LIMITS OF LIABILITY

Under no circumstances shall LaMotte Company be liable for loss of life, property, profits, or other damages incurred

through the use or misuse of their products.

PACKAGING AND DELIVERY

Experienced packaging personnel at LaMotte Company assure the adequate protection against normal hazards encountered

in transportation of shipments. After the product leaves the manufacturer, all responsibility for its safe delivery is assured by

the transportation company. Damage claims must be filed immediately with the transportation company to receive

compensation for damaged goods.
Should it be necessary to return the instrument for repair or servicing, pack instrument carefully in suitable container with

adequate packing material. Attach a letter to the shipping carton describing the kind of trouble experienced. This valuable

information will enable the service department to make the required repairs more efficiently.

EPA COMPLIANCE

The DC1600 Colorimeter is an EPA-Accepted instrument. EPA-Accepted means that the instrument meets the

requirements for colorimeters as found in test procedures that are approved for the National Primary Drinking Water

Regulations (NPDWR) or National Pollutant Discharge Elimination System (NPDES) compliance monitoring programs.

EPA-Accepted instruments may be used with approved test procedures without additional approval.

REPLACING LIGHT BULB

Turn the meter over, making sure the battery compartment is in the upper left corner (This is important). Remove the four

screws from the bottom of the colorimeter and remove the base. The burned out light bulb is attached to the small

rectangular circuit located just to the right (your right) of the light chamber. Remove the two screws that connect the

circuit and SAVE THE BURNED OUT LIGHT BULB. The light bulb must be returned to LaMotte Company for

replacement. Make sure the two washers are still in place. Remove the screw in the upper left corner of the colorimeter and

detach the replacement circuit. Replace that screw. When fastening the fresh bulb in place, be sure both washers are

aligned. Align the base to the meter and replace the four original screws.

NOTE:

If the re place ment bulb is sig nif i cantly dif fer ent from the orig i nal bulb, the “Set Blank” con trol may not have

enough range; if so, please call our tech ni cal sup port peo ple for as sis tance.

REPLACING THE BATTERY

The colorimeter is equipped with a battery check indication, the symbols BAT and ~ on the left hand side of the display,

that will be displayed when the battery needs to be replaced. The meter will still provide valid readings, but the readings

may drift. Eventually the meter will not have enough power to turn on. To replace the battery, remove the panel on the

back of the meter and detach the battery. Replace with a fresh alkaline 1604A type (9V) battery.

Battery polarity (+ & –) must never be reversed, even momentarily. If it is, the instrument will be rendered

INOPERABLE, and must be returned to LaMotte Company for repair. This will be considered a non-warranty repair.

Use appropriate caution when replacing the battery.