LaMotte DC1100 Colorimeter User Manual
Page 8
TROUBLE SHOOTING GUIDE (continued)
PROBLEM
CHECK
ACTION
Suspect
Calibration
Verify standard with
alternate test method?
Run alternate test on standard.
^
Verify with another
meter?
Check other meter calibrations.
^
Contact LaMotte for
return authorization.
Return for Calibration check.
^
Meter won’t zero
Sample tubes free of dirt
& scratches?
Check & replace if necessary.
^
Sample tubes clean?
Clean tubes often with soap and a test
tube brush. Do not leave reacted
samples in tubes, staining may occur.
^
Possible internal meter
components subjected to
moisture?
Always dry tubes before inserting.
Examine tube chamber for visible
moisture or chemical stains. If yes,
remove 4 screws from back, split case &
dry with hair dryer under low heat (10 -
15 min).
^
LED contaminated with
dirt and deposits?
Clean square LED and Photodetector
w/cotton swab & small amount of
alcohol.
^
High suspended solids in
sample blank?
Filter with coffee filter, or similar, to
remove most solids. (Be sure to filter
reacted samples as well)
NOTE: Do not use filtered blank for
sample reaction.
Meter read 1 -
w/o samples in
chamber
Meter over range?
Contact LaMotte for return
authorization & return.
Meter read 1 -
only when samples
in chamber
Meter over range?
Extremely high concentration, dilute
sample to see if 1 - condition returns to
normal condition.
12/95