TROUBLE SHOOTING GUIDE (continued)

PROBLEM

CHECK

ACTION

Suspect
Calibration

Verify standard with
alternate test method?

Run alternate test on standard.

^

Verify with another
meter?

Check other meter calibrations.

^

Contact LaMotte for
return authorization.

Return for Calibration check.

^

Meter won’t zero

Sample tubes free of dirt
& scratches?

Check & replace if necessary.

^

Sample tubes clean?

Clean tubes often with soap and a test
tube brush. Do not leave reacted
samples in tubes, staining may occur.

^

Possible internal meter
components subjected to
moisture?

Always dry tubes before inserting.
Examine tube chamber for visible
moisture or chemical stains. If yes,
remove 4 screws from back, split case &
dry with hair dryer under low heat (10 -
15 min).

^

LED contaminated with
dirt and deposits?

Clean square LED and Photodetector
w/cotton swab & small amount of
alcohol.

^

High suspended solids in
sample blank?

Filter with coffee filter, or similar, to
remove most solids. (Be sure to filter
reacted samples as well)
NOTE: Do not use filtered blank for
sample reaction.

Meter read 1 -
w/o samples in
chamber

Meter over range?

Contact LaMotte for return
authorization & return.

Meter read 1 -
only when samples
in chamber

Meter over range?

Extremely high concentration, dilute
sample to see if 1 - condition returns to
normal condition.

12/95