TROUBLE SHOOTING GUIDE

PROBLEM

CHECK

ACTION

Meter won’t
come on

Battery?

Replace, see pg. 6 of manual.

^

AC Adapter?

See pg. 6 of manual.

^

AC Wall Outlet?

Verify power source.

^

Contact LaMotte for
return authorization.

Return to LaMotte for repair.

^

Comes on, but
readings drift

Are instrument grade
reagents being used?

Obtain instrument grade reagents.

^

Are there suspended solids
in sample?

Filter both blank and sample
(after reaction).
NOTE: Do not use filtered blank for
sample reaction.

^

Do reading drift up scale? Bring samples to room temperature.

^

Readings drift down scale
with cold sample?

Check for sample tube fogging, let
sample equilibrate to room temperature.

Suspect
Calibration

Will meter blank? (zero)

If no, see next page.

^

Are instrument grade
reagents being used?

Obtain instrument grade reagents.

^

How old are the reagents? Contact LaMotte to verify age.

^

Check calibration with
standards?

How old is the standard?

^

Is class “A” volumetric
glassware being used to
make up the standards?

If not, obtain and see Standard
Methods.