Appendix: electronic flow control option, Technical data for mass flow controllers – VICI 230 Dynacalibrator User Manual
Page 39

Appendix: Electronic Flow Control Option
Technical Data for Mass Flow Controllers
Specification
Mass Controller
Description
Accuracy
± (0.8% of reading + 0.2% of full scale)
At calibration conditions after tare
High accuracy option
± (0.4% of reading + 0.2% of full scale)
At calibration conditions after tare
Repeatability
± 0.2%
Full scale
Operating range
1/2% to 100%
Full scale
Turndown ratio
200:1
Typical response time
100
Milliseconds (adjustable)
Standard conditions (STP)
25°C and 14.696PSIA
Mass reference conditions
Operating temperature
-10 to +50
°Celsius
Zero shift
0.02%
Full scale/°Celsius/Atm
Span shift
0.02%
Full scale/°Celsius/Atm
Humidity range
0 to 100%
Non-condensing
Controllable flow rate
102.4%
Full scale
Maximum pressure
145
PSIG
Input/Output signal digital
Mass flow, volumetric flow, pressure,
and temperature
RS-232 or RS-485 serial or
PROFIBUS or DeviceNet
Input/Output signal analog
Mass flow
0-5 VDC
Optional Input/Output signal
secondary analog
Mass flow, volumetric flow, pressure, or
temperature
0-5 VDC or 0-10 VDC or 4-20 mA
Electrical connections
8 pin mini-DIN, DB9, or DB15
Supply voltage
12 to 30 VDC
(15-30 VDC for 4-20 mA outputs)
Supply current
0.250 amp (at 12 VDC, declining with
increased supply voltage)
Mounting attitude sensitivity
None
Warm-up time
< 1
Second
Wetted materials
303 and 302 stainless steel, Viton®, silicone RTV (rubber), glass reinforced
nylon, aluminum, brass, 430FR stainless steel, silicon, glass