Entering commands – VICI 230 Dynacalibrator User Manual
Page 24

0
10. Click “ASCII Setup” and set the parameters as indicated below.
Click “OK”.
11. When the screen returns to Dyna230 HyperTerminal (page 18, bottom illustra-
tion), click “File”, then “Save” to save this Dynacalibrator Model 230 communi-
cations setup.
The following text will be displayed in the HyperTerminal window when
communication is first established with the Dynacalibrator Model 230:
Parameters recovered.
Program checksums ok.
Initializing SPI . . . ok
Initializing A/D . . . ok
Initializing IIC . . . DS1624 started
Valco Dynacalibrator online.
A simple processor test can be performed by opening HyperTerminal before
powering up the Dynacalibrator. If the processor has some defects, the message
“Bad checksum in Block xxx” is displayed instead of line 2 above. If this occurs,
please consult VICI.
Entering Commands
Each command must be followed by the
Examples:
To set the chamber temperature to 50°C, type “TS=50” and press
To read the set point, type “TS” and press
To read the current chamber temperature, type “TR” and press
To start data logging, type “L+” and press
logging, type “L-” and press
Serial Port Communication