PART X : UPDATING FIRMWARE

10.1 Hardware Requirements

10.2 Preparation

10.3 Updating Firmware

1.

A SynScan hand control with firmware version 3.0 or above.

2.

A computer running Windows 95 or a later version.

3.

A RS-232C serial port on the computer, or a USB-to-Serial adaptor.

4.

The PC-Link cable (D-Sub9 to RJ-12) which comes with the mount.

5.

A DC 7.5V to 12V DC power supply which can output at least 100 mA current.

1.

Create a new folder in the computer (for example,

C:\SynScan

) to save relevant files.

2.

Download the application package “

SynScan Firmware Loader

” from http://www.skywatch-

er.com/ and extract the file “

SynScanFirmwareLoader.exe

” to the above folder.

3.

Download the latest firmware package and extract the “

.SSF

” file to the above folder.

1.

Connect the computer and the SynScan hand control with the PC-Link cable.

2.

Press and hold the “

0

” and “

8

” keys simultaneously, and then power on the hand control.

The hand control will give a beep, and display “

SynScan Update

” on the screen. Release

the “

0

” and “

8

” keys.

3.

On the computer, run the

SynScanFirmwareLoader.exe

. An application window is show in

Fig 10.3a.

• Use the “

Browse

” button to load the latest firmware file (“

.SSF

” file)

• Check the “

Enforce database update

” to enforce updating the hand control’s database.

Clear it to let the application determine whether it is necessary to update the database.

Fig. 10.3a