Metrohm NIRS XDS Interactance OptiProbe Analyzer User Manual
Page 51
▪▪▪▪▪▪▪
49
Instrument Calibration uses a stable rare-
earth standard, of known wavelength
response, as a method to establish
wavelength scale response of the instrument.
This standard is directly traceable to NIST
SRM-1920a.
The instrument is set to scan the standard,
and the nominal peak positions for each
major absorption are determined. Vision
performs an algorithm to set the peak
positions of the instrument to those of the
standard. These wavelength positions are
applied at the sample plane, where samples
are measured or read.
These adjustments are saved, and are applied
on each subsequent scan of the instrument,
yielding a correct spectrum.
This adjustment is performed in reflectance,
using the Interactance Reflectance Probe tip.
Once done, this calibration applies to
Interactance Reflectance and Interactance
Immersion
Vision takes an instrument reference, which
takes about 20 seconds. The Interactance
Reflectance Probe must be in the uppermost
position in the orange holder.
This dialog box is displayed.
- 915 KF Ti-Touch (382 pages)
- 800 Dosino (53 pages)
- 767 Calibrated Reference (23 pages)
- 940 Professional IC Vario ONE/SeS/Prep 2 (54 pages)
- 754 Dialysis Unit (49 pages)
- 815 Robotic Soliprep for LC (76 pages)
- Vision Manual (207 pages)
- tiamo 2.1 Manual (1532 pages)
- 825 Lab Link (37 pages)
- 808 Titrando (70 pages)
- 902 Titrando (52 pages)
- 756 KF Coulometer (163 pages)
- 756 KF Coulometer (162 pages)
- 940 Professional IC Vario ONE/LPG (98 pages)
- 850 Professional IC Anion MCS Prep 3 (152 pages)
- 850 Professional IC Anion MCS Prep 3 (154 pages)
- 904 Titrando (58 pages)
- 850 Professional IC Anion MSM-HC MCS Prep 2 (150 pages)
- 930 Compact IC Flex Oven/ChS/Deg (47 pages)
- 872 Extension Module Liquid handling (64 pages)
- 814 USB Sample Processor (91 pages)
- 814 USB Sample Processor (90 pages)
- 940 Professional IC Vario (43 pages)
- Vision – Tutorial (40 pages)
- 799 GPT Titrino (242 pages)
- 889 IC Sample Center (68 pages)
- 761 Compact IC (228 pages)
- 851 Titrando (100 pages)
- 748 DH Sample Changer (32 pages)
- 940 Professional IC Vario ONE/SeS/HPG (51 pages)
- 896 Professional Detector – Amperometry (62 pages)
- 877 Titrino plus (139 pages)
- 881 Compact IC pro – Anion (129 pages)
- 940 Professional IC Vario ONE/ChS/HPG (112 pages)
- 930 Compact IC Flex Deg (41 pages)
- 840 PC Control 5.0 / Touch Control (351 pages)
- 940 Professional IC Vario ONE/Prep 1 (45 pages)
- 776 Dosimat (42 pages)
- 717 Sample Changer (36 pages)
- 815 Robotic USB Sample Processor XL (114 pages)
- 815 Robotic USB Sample Processor XL (113 pages)
- 940 Professional IC Vario ONE/SeS/PP (126 pages)
- 838 Advanced Sample Processor Installation Instructions (109 pages)
- 700 Dosino (55 pages)
- 719 S Titrino (152 pages)