Application Bulletin 287 e
Installation Instructions of TitrIC 3
Page 5/16

3 Cable

connections

The numbers given here (

) refer to the drawing on the next page.

First all the power cables are connected to the instruments. Both the instruments and the PC remain
switched off.

USB cable from Edgeport to PC – supplied with USB Converter Edgeport/4 (2.145.0320)

RS cable (6.2125.110) from 712 Conductometer to Edgeport port 1

RS cable (6.2134.100) from 2.861.0020 IC to Edgeport port 2 – supplied with 861

RS cable (6.2134.100) from 2.861.0010 IC to Edgeport port 3 – supplied with 861

USB cable (6.2151.000) from PC to 809 Controller, do not connect!

MSB cable from Dosino 800, 50 mL (anions) to 809 MSB 1, do not connect!

MSB cable from Dosino 800, 10 mL (cations) to 809 MSB 2, do not connect!

MSB cable from Dosino 800, 20 mL (0.1 M HCl) to 809 MSB 3, do not connect!

USB cable (6.2151.000) from 809 to 815 Controller – supplied with 815, do not connect!