2.1. Special messages and error messages

775 Dosimat

4

2 Error messages, troubleshooting

2.1 Special messages and error messages

error 1

Check sum error in PROM.

Remedy: Call Metrohm-Service.

error 2

RAM-check: Error in on-chip-RAM.

Remedy: Call Metrohm-Service.

error 3

RAM-check: Error in off-chip-RAM.

Remedy: Call Metrohm-Service.

error 4

RAM-check: Error in on- and off-Chip-RAM.

Remedy: Call Metrohm-Service.

error 5

Check sum error in off-chip-RAM.

Remedy: RAM has to be re-initialized: Switch Dosimat off.

Press during switching it on again. Display shows

"RAM init.". Press . Display shows "RAM init. passed".

leads to basic program.

no exch. unit!

Exchange unit is not (properly) mounted.

Exit: Mount Exchange unit properly.