Page 14

background image

3.1. Technical specifications

775 Dosimat

10

3 Appendix

3.1 Technical specifications

Exchange units

1, 5, 10, 20, 50 mL burette cylinder volumes,

preferably with flat cock for automatic cock changeover

Resolution

10'000 pulses per 100% of burette volume

Exactitude

Metrohm dosimats and exchange units meet the requirements

of ISO/EN/DIN Standard 8655-3 "Piston-operated volumetric

apparatus – Part 3: Piston burets" and DIN Standard 12 650.

Limits according to ISO/EN/DIN 8655-3

Cylinder

volume

Max. permissible system. error

Max. permissible random error

1 mL

± 0.6%

± 6 µL

± 0.1%

± 1 µL

5 mL

± 0.3%

± 15 µL

± 0.1%

± 5 µL

10 mL

± 0.3%

± 20 µL

± 0.07%

± 7 µL

20 mL

± 0.2%

± 40 µL

± 0.07%

± 14 µL

50 mL

± 0.2%

± 100 µL

± 0.05%

± 25 µL

Metrohm agencies throughout the world offer you the possibil-

ity of checking the accuracy of your exchange units and Dosi-

mats locally and also of certifying them. If the dosing cylinder

and/or piston of an exchange unit are replaced then we rec-

ommend that an accuracy check is carried out.

Dispensing time for 100% of burette cylinder volume

20 s ... app. 17 min

Display

LCD, 16 characters

Size of characters: 4.84 x 8.01 mm

Material

Cabinet

Polybutylene terephthalate (PBTP)

Key cover

Polycarbonate (PC)

Ambient temperature Nominal functional range

+5... +40°C

Storage, transport

- 40... +60°C

Safety specifications

Designed and tested in accordance to IEC-Publication 1010,

safety class I. This manual contains some information and

warnings which have to be followed by the user to ensure safe

operation and to retain apparatus in safe condition.