4 appendix, 1 technical data, Warranty – Metrohm 782 IC Column Thermostat User Manual
Page 18: Eu declaration of conformity, Certificate of conformity and system validation, 4appendix
4.1 Technical data
782 IC Column Thermostat
15
4
Appendix
4.1
Technical data
Temperature control
Temperature range
30
…
75 °C, to be set in steps of 0.1 °C
Accuracy
±
0.5 °C
Stability
±
0.2 °C
Mains connection
Mains voltage
115 V: 100...120 V
±
10 %
230 V: 220...240 V
±
10 %
Switchable with voltage selection insert in fuse
holder (see section 2.2.1)
Mains frequency
50...60 Hz
Power consumption
12 VA
Fuse
5 mm Ø, 20 mm length
100
…
120 V: 0.5 A (slow-blow)
220
…
240 V: 0.25 A (slow-blow)
Safety specifications
Construction / Testing
According to IEC 1010 / EN 61010 / UL 3101-1,
protection class 1, degree of protection IP40
Safety directions
The Instructions for Use include information and
warnings to which the user must pay attention in
order to assure safe operation of the instrument.
Electromagnetic compatibility (EMC)
General requirements
Standards met:
IEC61326/EN61326, NAMUR
Emitted interference
Standards met:
EN55011, EN55022, EN 50081-1/2
Immunity to interference
Standards met:
IEC61000-4-2/EN61000-4-2 (class 4),
IEC61000-4-4/EN1000-4-4 (class 4),
IEC61000-4-11/EN61000-4-11,
IEC61000-4-14/EN61000-4-14,
IEC61000-6-2/EN61000-6-2,
EN50082-1/2