2 Installation

782 IC Column Thermostat

12

capillary to the inlet end of the separating column 81

81 (note

flow direction).

4 Rinse column

Place a beaker beneath the column outlet.

Open software window for manual system control.

If necessary, modify

Flow rate

to the value suited for the

inserted separting column and click on

to send

this value to the 761 Compact IC.

Switch on high-pressure pump (

IC pump

) by clicking

and

rinse column with eluent for ca. 10 min.

Switch off high-pressure pump by clicking

.

5 Connect column to detector block or suppressor

module

without suppressor module:

Screw outlet end of separating column 81

81 to the inlet capillary

45

45 permanently mounted on the detector block 46

46.

with suppressor module:

Screw outlet end of separating column 81

81 to suppressor inlet

capillary 96

96 using a 6.2744.010 compression fitting.

6 Insert column in IC Column Heater

Remove cover of IC Column Heater.

Insert separating column with inlet and outlet capillaries into

the IC Column Heater in such a way that the outlet end of the

separting column is at the connecting cable side.

Insert inlet and outlet capillaries in the slices made for them

and close IC Column Heater with the cover.

7 Fix IC Column Heater

Insert the two 6.2027.060 holders in the mounting rails 27

27

(see Fig. 14 of 761 Instruction for Use) and fasten IC Column

Heater in the holder so that the connecting cable is on the

upper side.

8 Connect IC Column Heater

Plug the connecting cable permanently attached to the IC

Column Heater into connection 11

11 "IC Column Heater“ of the

782 IC Column Thermostat (see Fig. 2).