Metrohm 782 IC Column Thermostat User Manual
Page 15
2 Installation
782 IC Column Thermostat
12
capillary to the inlet end of the separating column 81
81 (note
flow direction).
4 Rinse column
•
Place a beaker beneath the column outlet.
•
Open software window for manual system control.
•
If necessary, modify
Flow rate
to the value suited for the
inserted separting column and click on
to send
this value to the 761 Compact IC.
•
Switch on high-pressure pump (
IC pump
) by clicking
and
rinse column with eluent for ca. 10 min.
•
Switch off high-pressure pump by clicking
.
5 Connect column to detector block or suppressor
module
•
without suppressor module:
Screw outlet end of separating column 81
81 to the inlet capillary
45
45 permanently mounted on the detector block 46
46.
•
with suppressor module:
Screw outlet end of separating column 81
81 to suppressor inlet
capillary 96
96 using a 6.2744.010 compression fitting.
6 Insert column in IC Column Heater
•
Remove cover of IC Column Heater.
•
Insert separating column with inlet and outlet capillaries into
the IC Column Heater in such a way that the outlet end of the
separting column is at the connecting cable side.
•
Insert inlet and outlet capillaries in the slices made for them
and close IC Column Heater with the cover.
7 Fix IC Column Heater
•
Insert the two 6.2027.060 holders in the mounting rails 27
27
(see Fig. 14 of 761 Instruction for Use) and fasten IC Column
Heater in the holder so that the connecting cable is on the
upper side.
8 Connect IC Column Heater
•
Plug the connecting cable permanently attached to the IC
Column Heater into connection 11
11 "IC Column Heater“ of the
782 IC Column Thermostat (see Fig. 2).