2 pump tubing connector with filter, Figure 13, Pump tubing connection – changing the filter – Metrohm 872 Extension Module Sample Prep User Manual
Page 31

■■■■■■■■■■■■■■■■■■■■■■
5 Handling and maintenance
872 Extension Modul – Sample Prep
■■■■■■■■
23
Order
number
Name
Material
Inner
diameter
Use
6.1826.360
Pump tubing LFL
(white/white), 3-stop-
per
PVC (Tygon)
1.02 mm
For sample transfer.
6.1826.380
Pump tubing LFL
(gray/gray), 3-stopper
PVC (Tygon)
1.25 mm
For inline sample dilution.
6.1826.390
Pump tubing LFL (yel-
low/yellow), 3-stop-
per
PVC (Tygon)
1.37 mm
For the sample solution in
inline ultrafiltration.
5.2.2.2
Pump tubing connector with filter
The 6.2821.130 filters (13-2) should be changed every 3 months, more
frequently at higher backpressure.
1
2
3
Figure 13
Pump tubing connection – changing the filter
1
Tubing olive
2
Filter 6.2821.130
Packaging contains 10 items.
3
Filter screw
Replacing the filter
1 Unscrewing the filter screw
■
Unscrew the filter screw (13-3) out of the tubing olive (13-1).
2 Replacing the filter
■
Remove old filter (13-2).
■
Place new filter (13-2) flat in the tubing olive (13-1).
3 Mounting the filter screw
■
Screw filter screw (13-3) back into the tubing olive (13-1).