3 temperature controller, 4 sample holder for standard vials, Temperature controller – Metrohm 832 KF Thermoprep User Manual
Page 9: Sample holder for standard vials, Fig. 3, Fig. 4, Sample holder
1.3 Parts and controls
832 KF Thermoprep, Introduction
5
1.3.3 Temperature controller
Fig. 3 Temperature controller
31
Actual value display
red, 4-place
35
Output status display
yellow, for outputs 1 to 3
(K1 and K3: oven heating, K2: fan)
32
Set value display
green, 4-place
36
LED for ramp/program function
green, lights if configured
33
Program key
for selecting the parameters
37
EXIT key
to exit the dialog levels
34 Decrement key
for changing values
38
Increment key
for changing values
1.3.4 Sample holder for standard vials
Fig. 4 Sample holder
The dimensions of the 6.2063.000 sample holder are optimized for the
Metrohm 6.2419.007 standard sample vials (6 mL). This ensures the
highest possible heat flow from the oven to the sample.
If other sample containers with different dimensions shall be used
custom made sample holders can be delivered. For this the precise
dimensions (incl. dimensional tolerances) must be known.
The measures of non-standard sample containers must vary within the
following dimensional limits:
Diameter 10.0 … 32.0 mm
Depth of immersion 20.0 … 45.0 mm
31
32
35
36
33
37
34
38