5.1 Problems and their solutions
■■■■■■■■■■■■■■■■■■■■■■
88
■■■■■■■■
850 Professional IC –Cation – Prep 1
Problem
Cause
Remedy
Sample – Blockage in the
sample path.
Check the sample path.
Sample – Sample loop not
(completely) filled.
Prolong the sample transfer time.
Sample – Gas bubbles in
the sample.
Use the sample degasser (see Chapter 2.13,
page 42) (if available).
Peristaltic pump –
insufficient or no
delivery rate
Peristaltic pump – contact
pressure too weak.
Correctly set contact pressure (see "Set flow
rate", page 47).
Peristaltic pump – filter
blocked.
Replace the filter (see Chapter 4.10.2.2, page
83).
Peristaltic pump – pump
tubing defective.
Replace pump tubing (see Chapter 4.10.2.1,
page 82).
Very noisy baseline
High pressure pump – con-
taminated pump valves.
Clean pump valves (see Chapter 4.5.2, page
67).
Eluent – Leak in the eluent
path.
Check the eluent path.
Eluent – Blockage in the
eluent path.
Check the eluent path.
High pressure pump –
defective piston seals.
Replace piston seals (see Chapter 4.5.2, page
67).
Pulsation damper not con-
nected.
Connect the pulsation damper (see Chapter
2.12, page 40).
Pulsation damper not con-
nected.
Connect pulsation damper (see Chapter 2.12,
page 40).
Background conduc-
tivity too high
Incorrect eluent.
Change eluent (see Chapter 4.4.2.3, page 66).
Data of the separa-
tion column cannot
be read.
Column chip contamina-
ted.
Clean contact surfaces of the column chip
(with alcohol).
Column chip defective.
1. Save column configuration in MagIC Net.
2. Notify Metrohm Service.
Individual peaks
greater than expec-
ted
Sample – Carry-over of the
samples from previous
measurement.
Rinse system longer between two samples.