5 troubleshooting, 1 problems and their solutions, Problems and their solutions – Metrohm 850 Professiona IC Cation Prep 1 User Manual
Page 97

■■■■■■■■■■■■■■■■■■■■■■
5 Troubleshooting
850 Professional IC –Cation – Prep 1
■■■■■■■■
87
5 Troubleshooting
5.1
Problems and their solutions
Problem
Cause
Remedy
Marked drop in
pressure
Leak in the system.
Check and seal connections (see Chapter 2.5,
page 16).
Marked rise in pres-
sure
6.2821.120 inline filter
blocked.
Replace 6.2821.130 filter (see Chapter 4.6,
page 77).
Conductivity detector
blocked.
■
Shorten capillary ends by several mm (see
Chapter 4.12.1, page 84).
■
Rinse detector against the normal flow
direction (see Chapter 4.12.1, page 84).
Guard column – blocked.
Replace guard column (see Chapter 2.19,
page 56).
Separation column –
blocked.
■
Regenerate separation column (see Chap-
ter 4.13.4, page 86).
■
Replace separation column (see "Connect-
ing and rinsing the separation column",
page 59).
Note: Samples should always be microfiltered
(see Chapter 4.7, page 79).
Injection valve – valve
blocked.
Have the valve cleaned (by Metrohm service
technicians).
Drift of the baseline
Thermal equilibrium not yet
attained.
Condition (see Chapter 2.16, page 50) instru-
ment with the column thermostat (see Chap-
ter 3.2, page 61) switched on .
Leak in the system.
Check and seal connections (see Chapter 2.5,
page 16).
Eluent – Evaporation of the
organic solvent in the elu-
ent.
Check the eluent bottle attachment (see Figure
14, page 31).
Peak areas lower
than expected
Sample – Leak in the sam-
ple path.
Check the sample path.