beautypg.com

Metrohm 771 IC Compact Interface User Manual

Page 24

background image

4 Appendix

771 IC Compact Interface

18

Noise

< 20

µ

Vpp

< 3

µ

Vrms (at 0 V input voltage, amplification 1

and sampling rate 10 Hz)

Zero error

±

2.5 mV

Pin assignment

Pin 1,4,5:

GND (ground, connected to
housing)

Pin 2:

Analog +

Pin 3:

Analog –

Pin 8:

RUN

Pin 9:

COM

Safety specifications

Construction/testing

According to IEC 1010 / EN 61010 / UL 3101-1,

protection class 1, degree of protection IP40

Safety directions

The Instructions for Use include information and

warnings which must be heeded by the user to

assure safe operation of the instrument.

Electromagnetic compatibility (EMC

Emitted interference

Standards met:

EN55011 (level B), EN55022 (level B),

EN 50081-1/2, EN61000-3-2

Immunity to interference

Standards met:

IEC61000-4-2/EN61000-4-2 (level 3),

IEC61000-4-3/EN61000-4-3 (level 3),

IEC61000-4-4/EN61000-4-4 (level 4),

IEC61000-4-5/EN61000-4-5 (level 2/3),

IEC61000-4-6/EN61000-4-6 (level 3),

IEC61000-4-11/EN61000-4-11,

IEC61000-4-14/EN61000-4-14 (level 3),

IEC61000-6-2, IEC61326-1+A1,

EN50082-1, NAMUR

Ambient temperature

Nominal operating range

+5

+45°C

(at 20

80 % atmospheric humidity)

Storage, transport

–40

+70°C

Housing

Material

Steel, enameled

Dimensions

Width

183 mm

Height

79 mm

Depth

165 mm

Weight

1.9 kg (without accessories)