Metrohm 771 IC Compact Interface User Manual
Page 24
![background image](/manuals/586741/24/background.png)
4 Appendix
771 IC Compact Interface
18
Noise
< 20
µ
Vpp
< 3
µ
Vrms (at 0 V input voltage, amplification 1
and sampling rate 10 Hz)
Zero error
±
2.5 mV
Pin assignment
Pin 1,4,5:
GND (ground, connected to
housing)
Pin 2:
Analog +
Pin 3:
Analog –
Pin 8:
RUN
Pin 9:
COM
Safety specifications
Construction/testing
According to IEC 1010 / EN 61010 / UL 3101-1,
protection class 1, degree of protection IP40
Safety directions
The Instructions for Use include information and
warnings which must be heeded by the user to
assure safe operation of the instrument.
Electromagnetic compatibility (EMC
Emitted interference
Standards met:
EN55011 (level B), EN55022 (level B),
EN 50081-1/2, EN61000-3-2
Immunity to interference
Standards met:
IEC61000-4-2/EN61000-4-2 (level 3),
IEC61000-4-3/EN61000-4-3 (level 3),
IEC61000-4-4/EN61000-4-4 (level 4),
IEC61000-4-5/EN61000-4-5 (level 2/3),
IEC61000-4-6/EN61000-4-6 (level 3),
IEC61000-4-11/EN61000-4-11,
IEC61000-4-14/EN61000-4-14 (level 3),
IEC61000-6-2, IEC61326-1+A1,
EN50082-1, NAMUR
Ambient temperature
Nominal operating range
+5
…
+45°C
(at 20
…
80 % atmospheric humidity)
Storage, transport
–40
…
+70°C
Housing
Material
Steel, enameled
Dimensions
Width
183 mm
Height
79 mm
Depth
165 mm
Weight
1.9 kg (without accessories)