Metrohm 850 Professional IC Anion MSM-HC MCS Gradient User Manual
Page 28

3.3 Installation diagram
■■■■■■■■■■■■■■■■■■■■■■
18
■■■■■■■■
850 Professional IC
3
Connection capillary
Capillary of the MSM-HC labeled with Elu-
ent.
4
Connection capillary
Capillary of the MSM-HC labeled with
Detector.
5
Detector inlet capillary
6
Sample aspirating capillary 6.1803.040
Connected to the injection valve. The other
end can be connected either to the sample
degasser or guided directly out of the instru-
ment.
7
PTFE connection capillary
Section of the 6.1803.040 PTFE capillary.
Connects sample degasser with sample pro-
cessor (required only if the sample degasser
is to be used).
8
Sample outlet capillary
Connected to the injection valve.
9
Rinsing solution aspirating capillary
6.1803.020
Connection to the bottle with rinsing solu-
tion.
10 Pump tubing 6.1826.320
Pump tubing with yellow-orange stoppers
for conveying the rinsing solution.
11 Connection capillary for the rinsing
solution
Capillary of the MSM-HC labeled with H20.
12 Rinsing solution outlet capillary
Capillary of the MSM-HC labeled with
Waste.
13 Regeneration solution aspirating capil-
lary 6.1803.020
Connection to the regeneration solution
bottle.
14 Pump tubing 6.1826.320
Pump tubing with yellow-orange stoppers
for conveying the regeneration solution.
15 Connection capillary for the regenera-
tion solution
Capillary of the MSM-HC labeled with
H2SO4.
16 Regeneration solution outlet capillary
Capillary of the MSM-HC labeled with
Waste.
17 Air aspirating capillary
For aspirating CO
2
-low air from the car-
tridges of the MCS.
18 PEEK pressure screws, short
6.2744.070
19 PEEK pressure screws, long
6.2744.090
20 Tubing olive 6.2744.030
21 Pump tubing connection with safety
device 6.2744.180