3 Installation
850 Professional IC
■■■■■■■■
15
■
Connect the 6.1803.020 aspirating capillary for regeneration solu-
tions with a 6.2744.034 tubing olive and a 6.2744.070 short
pressure screw to the aspiration end of the 6.1826.320 pump
tubing for the regeneration solution.
■
Place the pump tubing in a tubing cartridge.
■
Connect the 6.1803.020 aspirating capillary for rinsing solutions
with a 6.2744.034 tubing olive and a 6.2744.070 short pressure
screw to the aspiration end of the second 6.1826.320 pump tub-
ing for the rinsing solution.
■
Place the second pump tubing in the second tubing cartridge.
■
Place both tubing cartridges into the peristaltic pump.
6 Connecting the MSM-HC
■
Connect the capillary labeled with H2SO4 with a 6.2744.180
pump tubing connection and a 6.2744.070 short pressure screw
to the peristaltic pump on the outlet end of the pump tubing for
the regeneration solution.
■
Connect the capillary labeled with H20 with a 6.2744.180 pump
tubing connection and a 6.2744.070 short pressure screw to the
peristaltic pump on the outlet end of the pump tubing for the
rinsing solution.
■
Guide the two capillaries labeled with Waste into a waste con-
tainer and fasten them there.
7 Connecting the instrument
■
Connecting the instrument to the PC with a 6.2151.020 USB
cable (see Chapter 3.20.1, page 66).
■
Connect instrument to mains supply (see Chapter 3.20, page
66).
8 Initial start-up
■
Switch on the PC and start MagIC Net.
■
Switch on the instrument.
■
Deaerate the high pressure pump (see Chapter 3.10.2, page
40).
■
Set contact pressure of the peristaltic pump (see "Set flow rate",
page 59).
■
Rinse instrument without column.