3.2 Installing the rinsing solution flow path
■■■■■■■■■■■■■■■■■■■■■■
8
■■■■■■■■
IC equipment: Liquid Handling Station
8 Inserting the adsorber tube
■
Insert the adapter into the SGJ opening of the bottle cap.
■
Screw the adsorber tube into the upper opening of the adapter.
Installing the capillary between the supply bottle and the
trap column
Required accessories
■
PEEK capillary (6.1831.180)
■
Metrosep I Trap 1 - 100/4.0 (6.1014.200)
■
Pressure screw (6.2744.010)
1 Removing the stoppers
■
Remove the stoppers from the trap column's inlet and outlet.
Keep the stoppers.
2 Installing the capillary to the column inlet
■
Tighten the other end of the installed capillary to the trap column
inlet using a pressure screw.