beautypg.com

Brightener determination, Manual operation, Workplace – Metrohm 894 Professional CVS semiautomated User Manual

Page 7: Course of events, Remarks, Storing dosing units, Storing electrodes, Waste bottle

background image

Application Bulletin 401/1 e

Installation instruction for 894 Professional CVS semiautomated

5.2.

Brightener determination

5.2.1. Manual operation

Dosing Unit «894/D1 50mL VMS» connected to Dosino

1 has to be prepared with VMS.

Dosing Unit «894/D2 2mL Brightener» connected to

Dosino 2 has to be prepared with brightener

concentrate.

Dosing Unit «894/D3 2mL Suppressor» connected to

Dosino 3 has to be prepared with suppressor

concentrate.

The correct «Sample amount» has to be pipetted into

the measuring vessel when requested.

5.2.2. Workplace

Select sample type S

AMPLE

in the «Run» window of the

«Workplace» to run a determination.

Define the volume of sample used for the determination

under «Sample amount» in the «Run» window of the

«Workplace».

5.2.3. Course of events

The execution of the determination by modified linear

approximation technique is controlled by the 894

Professional CVS and viva.

The intercept solution is automatically prepared at the

beginning of each determination, by dosing VMS from

Dosino 1 and suppressor concentrate from Dosino 3.

After the determination of the intercept value, a window

will pop up prompting to add the sample on top of the

intercept solution. The exact sample volume has to be

pipetted manually through the pipetting opening of the

894 Professional CVS.

Standard addition is carried out automatically with

brightener concentrate from Dosino 2.

6. Remarks

6.1.

Storing dosing units

When the dosing units are not used (during the night, over

the weekend) the dosing cylinder has to be rinsed with

water. Otherwise elemental copper can be formed or

additives can precipitate between the valve disk and the

distributor disk or in the tubings. A blockage can damage

the 800 Dosino. Emptying the cylinder and filling it with

water is sufficient to prevent a blockage.

6.2.

Storing electrodes

When the electrodes are not used (during the night, over the

weekend) the electrodes should be thoroughly rinsed.

Working and auxiliary electrode can either be stored in

deionized water or dry. The reference electrode should be

stored separately in a vial filled with deionized water (KCl

solution, in case KCl is used as bridge electrolyte), so that

the reference electrode immerses at least to the rim of the

electrolyte vessel.

6.3.

Waste bottle

The waste bottle must not be closed completely. For

pressure balancing in the waste bottle, keep at least one

opening unsealed. Overpressure would lead to a

malfunction of the 800 Dosinos.

Page 7 of 7