Installation instruction for 894 Professional CVS semiautomated

4.1.2. Dosing commands

Such as:

Define the volume of solution that should be dosed

when this command is due.

Such as:

Define the volume of solution that should be dosed

when this command is due.

Assign the Dosing Unit that should be used to run this

command.

Such as:

On the tab «General/Hardware» assign the 894

Professional CVS instrument to which the 800 Dosino is

connected.

4.2.

Evaluation

Settings regarding evaluation and documentation of the

determination are located in the «Evaluation» part of the

method. The templates already include all necessary

settings to determine brightener or suppressor concentration

in an acid copper bath. If modifications however should be

necessary, here is where important parameters are found:

4.2.1. Substances

In the «Substances» part settings for peak recognition and

baseline parameters are defined.

4.2.2. Standards

In the «Standards» part the concentration of the used

standard solution is defined.

4.2.3. Calibration

In the «Calibration» part the calibration method, such as DT

or MLAT, is defined as well as the regression type.

4.2.4. Results

In the «Results» part on the tab «Results» the substance is

selected for which the concentration should be calculated

and displayed.

On the tab «Database» the database is defined where

determinations should be stored. By default viva database

is used.

On the tab «Report» a report template for an automatic

printout can be defined. By default no printout is defined.

5. Operation

5.1.

Suppressor determination

5.1.1. Manual operation

Dosing Unit «894/D1 50mL VMS» connected to Dosino

1 has to be prepared with VMS.

Dosing Unit «894/D4 2mL Std or sample» connected to

Dosino 4 has to be prepared with suppressor standard

solution if a calibration is recorded or with plating bath

sample for the determination.

5.1.2. Workplace

Select sample type S

TANDARD

to run a calibration in the

«Run» window of the «Workplace». For a determination

sample type S

AMPLE

has to be selected.

No «Sample amount» needs to be defined in the «Run»

window on the «Workplace», since addition volumes for

the standard and the sample are defined in the method

commands ADD

STD and ADD

SAMPLE

DT.

5.1.3. Course of events

The execution of calibration or determination by dilution

titration is controlled by the 894 Professional CVS and

viva.

Dosino 1 automatically doses the VMS and Dosino 4 is

used for the automatic addition of suppressor standard

or plating bath sample.

Page 6 of 7