5 Troubleshooting

2.896.0010 896 Professional Detector – Conductivity

■■■■■■■■

21

5 Troubleshooting

5.1

Problems and their solutions

Problem

Cause

Remedy

Marked rise in pres-
sure

Conductivity detector is
blocked.

(see Chapter 4.2.2, page 19)

Shorten the capillary ends by several mm.

Rinse the detector against the normal flow
direction.

Conductivity detec-
tor is not recognized
in the software

No connection.

Check the connection of the detector
cable.

Switch the instrument off and on again
(after 15 seconds).