Application Bulletin 354
Installation Instruction of TitrIC pro II
Page 7 of 14
instruments. The injection valves are connected in
series and finally connected to the T-piece
(6.1808.060) at the robotic arm (using the screws
6.2744.080 and 6.2744.070). In this setup and
methods the following sequential arrangement was
implemented: T-piece – anion injection valve –
cation injection valve – 50 mL Dosino.
5. Miscellaneous
5.1. Membrane Pumps of the 815
The membrane pump 1 at tower 1 is used to supply the DI
water for rinsing of the external vessel; the membrane pump
2 is used to empty the external vessel.
Connect the membrane pump 2/tower 1 to the left connector
of the Distributor leading to the M8 aspiration tip for
emptying the external vessel. The upper part of the
membrane pump 2 is connected to the waste canister.
The upper part of membrane pump 1 is connected to the
right connector of the Distributor leading to the three M6
rinsing nozzles to rinse the external vessel. The lower part
of the membrane pump 1 is connected to the DI-water
canister.
5.2. Stirrer
The 802 Rod Stirrer is inserted through opening 9 of the
external cell, and connected to the tower 1 of the 815. The
propeller (6.1909.050) is pushed on from below.
Length of the aspiration tips in the external cell
Clamp the titration vessel to the titration vessel lid.
The M8 aspiration tip in opening 7 must reach right to
the base of the titration vessel, as this is the only way to
ensure that the external cell is cleaned properly.
The 3-way stopper with antidiffusion valve in opening 6
should reach the middle of the titration vessel.
5.3. Electrodes
The pH electrode (iAquatrode Plus with Pt 1000;
6.0277.300) is inserted into opening 4. The cable is
connected with the 854 iConnect (provided with the
905) to the 905 Titrando at the rear on measuring input
1.
If the electrode is not used for some time then it should
be returned to its holder and kept in the storage solution
(6.2323.000).
The electrode has to be filled up regularly with the
electrolyte solution (6.2308.020).
Conductivity cell (6.0915.100): The measuring cell
cable is connected to the opening marked ‘
Cond. Cell
’ of
the 856 Conductometer. The measuring cell is inserted
into the flow-through cell (6.2763.000). The flow-
through cell itself is mounted on the rod of the stand
support (6.2001.070).