Metrohm TitrIC pro II User Manual
Page 12
Application Bulletin 354
Installation Instruction of TitrIC pro II
Page 12 of 14
In tiamo
TM
it is typically used as a single determination
in autorepeat mode. In other words, it is always ready
to do a titration as soon MagIC requires it. To set it into
autorepeat go to the Run window, right-mouse-click and
select
Properties Run window
. Under Process select
Autorepeat determination
. Close the window with
.
TitrIC pro II_pH calibration manual:
Here the iAquatrode
is taken out of the titration vessel and placed into the
buffer solutions. Start this method only from tiamo
TM
!
TitrIC pro II_titer determination of HCl:
This method is
used for the titer determination. It has to be used in
combination with the MagIC Net method
TitrIC pro
II_titer determination of HCl
.
TitrIC pro II_conductivity calibration:
This method is used
for calibration of the conductivity sensor. It has to be
used in combination with the MagIC Net method
TitrIC
pro II_calibration conductivity
.
With the TitrIC pro – Combo Kit Dilution following analysis
can be carried out using the TitrIC pro II_dilution methods in
MagIC respectively tiamo
TM
:
The whole procedure is controlled by the MagIC Net
software which is the master software. In tiamo
TM
the
method
TitrIC pro II_dilution
is selected as a single
determination. Start the determination in autorepeat mode.
To set it into autorepeat go to the Run window, right-mouse-
click and select
Properties Run window
. Under Process
select
Autorepeat determination
. Close the window with
. Start the determination. Now tiamo
TM
is waiting for the
start signal from MagIC Net via RS232.
In the MagIC Net software select the method
TitrIC pro
II_dilution
. Enter the required sample position, sample
identification etc. into MagIC Net. For value 1 choose
according to the list below:
Value 1 = 1 complete analysis including titration and
ion chromatography.
Value 1 = 2 only an ion chromatography analysis (e.g.
a calibration)
Value 1 = 3 only a titration analysis
With Value 1 = 1 a complete analysis is carried out.
Depending on the entered dilution factor the following will be
done:
Dilution Factor = 1:
Depending on the measured conductivity an appropriate
dilution factor is automatically applied by MagIC Net. The
calculation factor is calculated according to Time Program
No. 2. The finally used dilution factor is the integer of the
conductivity divided by 250.
Dilution Factor >1:
If a dilution factor is known it can be directly entered into the
sample table. It will be diluted accordingly. The conductivity
measurement has no influence on the dilution in this case.
E.g. With Value 1 = 2 an ion chromatographic analysis will
be carried out with the entered dilution factor. No dilution
factor will be calculated.