Metrohm TitrIC pro II User Manual
Page 14

Application Bulletin 354
Installation Instruction of TitrIC pro II
Page 14 of 14
TitrIC pro II_conductivity calibration:
This method is used
for calibration of the conductivity sensor. It has to be
used in combination with the MagIC Net method
TitrIC
pro II_calibration conductivity
.
Import: Click on
Method
at the left (in MagIC Net and
tiamo
TM
), then select under
File / Method manager… / Edit /
Import
… Import the methods related to TitrIC pro II with the
MiPT kit from the MagIC Net CD. You will find the methods
in the folder example/TitrIC of the CD.
Note!
Alterations of these methods should only be carried out by a
person at the administrator level and who is thoroughly
familiar with tiamo
TM
and MagIC Net
With the TitrIC pro – Combo Kit Partial Loop setup the
following analysis can be carried out using the TitrIC pro
II_MiPT methods in MagIC respectively tiamo
TM
:
The whole procedure is controlled by the MagIC Net
software which is the master software. In tiamo
TM
the
method
TitrIC pro II_MiPT
is selected as a single
determination. Start the determination in autorepeat mode.
To set it into autorepeat go to the Run window, right-mouse-
click and select
Properties Run window
. Under Process
select
Autorepeat determination
. Close the window with
. Start the determination. Now tiamo
TM
is waiting for the
start signal from MagIC Net via RS232.
In the MagIC Net software select the method
TitrIC pro
II_MiPT
. Enter the required sample position, sample
identification etc. into MagIC Net. For value 1 choose
according to the list below:
Value 1 = 1 complete analysis including titration and
ion chromatography.
Value 1 = 2 only an ion chromatography analysis (e.g.
a calibration)
Value 1 = 3 only a titration analysis