7.10 Power connection
■■■■■■■■■■■■■■■■■■■■■■
88
■■■■■■■■
930 Compact IC Flex ChS (2.930.1200)
7.10
Power connection
Required voltage
100 - 240 V ± 10% (autosensing)
Required fre-
quency
50 - 60 Hz ± 3 Hz (autosensing)
Power consump-
tion
■
65 W for typical analysis application
■
25 W standby (conductivity detector to 40 °C)
Power supply unit
■
Up to 300 W maximum, electronically monitored
■
Internal fuse 3.15 A
7.11
Interfaces
USB
Input
1 USB upstream, type B (for connection to the PC)
Output
2 USB downstream, type A
MSB
2 MSB Mini DIN 8-pin (female) (for Dosino, stirrer, remote lines, etc.)
Detector
1 15-pin high density DSUB (female)
Column recogni-
tion
for an intelligent column
Leak sensor
1 jack plug
Further connec-
tions
Auxiliary
1 15-pin DSUB (female)
Service
1 15-pin DSUB (female)
7.12
Safety specifications
This instrument fulfills the following electrical safety requirements:
CE marking in accordance with the EU directives:
■
2006/95/EC (Low Voltage Directive, LVD)
■
2004/108/EC (EMC Directive, EMC)
Federal Inspectorate for Heavy Current Installations ESTI (Accreditation
Number SCESp 033)
■
Safety mark for certification type 2 in accordance with NEV (type
testing with market monitoring, EMC conformity)