8 injection valve, 9 column heater, Injection valve – Metrohm 881 Compact IC pro – Cation User Manual
Page 83: Column heater
■■■■■■■■■■■■■■■■■■■■■■
7 Technical specifications
881 Compact IC pro – Cation
■■■■■■■■
75
Pressure range
Pump
0…50.0 MPa (0…500 bar)
Pump head
0…35.0 MPa (0…350 bar) (applies for the standard PEEK pump head)
Residual pulsa-
tion
< 1 %
Safety shutdown
Function
Automatic shutdown upon reaching the pressure limit values
Maximum pres-
sure limit
■
Adjustable from 0.1…50 MPa (1…500 bar)
■
The pump is automatically shut down at the first piston stroke
above the maximum limit value
Minimum pres-
sure limit
■
Adjustable from 0…49 MPa (0…490 bar)
■
The shutdown mechanism is inactive at 0 MPa
■
The shutdown mechanism only becomes active 2 minutes after sys-
tem start
■
The pump is automatically shut down after 3 piston strokes below
the minimum pressure limit
7.8
Injection valve
Actuator time
typ.100 ms
Max. operating
pressure
35 MPa (350 bar)
Material
PEEK
7.9
Column heater
Type
Resistance heater for the thermostatization of one integrated column
with a length of up to 300 mm.
Adjustable tem-
perature range
+ 0…+ 80 °C, in increments of 0.1 °C
Heating
Ambient temperature + 5 °C … Ambient temperature + 40 °C
Temperature
reproducibility
± 0.2 °C
Stability
< 0.05 °C
Heating up time
< 30 minutes from 20 to 40 °C