5 interfaces, 1 rs 232 interface, Interfaces – Metrohm 818 IC Pump User Manual
Page 37: Rs 232 interface, Data transmission protocol, N 5 interfaces
5.1 RS 232 interface
818 IC Pump / 8.818.1023 Instructions for Use
31
5 Interfaces
5.1
RS 232 interface
5.1.1
Data transmission protocol
The RS 232 interface of the 818 IC Pump is configured as DTE (Data
Terminal Equipment) with the following technical specifications:
•
Standard
Data interface in accordance with EIA standard
RS 232C (DIN 66020, page 1)
•
Baud rate 9600
•
Data Bits
8
•
Stopp Bits
1
•
XON/XOFF
off
•
Control characters C
R
DEC 13
HEX 0D
L
F
DEC 10
HEX 0A
•
Max. line length
80 characters + C
R
L
F
•
Cable
For interconnections of the 818 IC Pump with
non-Metrohm units, only a shielded data cable
(e.g. Metrohm D.104.0201) may be used. The
cable shielding must be faultlessly earthed at
both units (pay attention to current loops;
always use star-head earthing). Only
connectors with adequate shielding may be
used (e.g. Metrohm K.210.0001 with
K.210.9004).
•
Cable length
max. ca. 15 m