Metrohm 818 IC Pump User Manual
Page 27

3.3 Settings in the "818 IC Pump" window
818 IC Pump / 8.818.1023 Instructions for Use
21
ENABLED
Enable program start (a disabled program is
not started).
Add new program instruction.
Delete selected program instruction.
Test the time program (error messages are
displayed if program is wrong).
List of program instructions
The following program instructions can be added to the time program
on the
Program
page:
Befehl Parametereintrag Bedeutung
Pump on,
off
Switch the high pressure pump on or
off.
Flow
0.01 ... 5 mL/min
Set flow rate of the high pressure
pump to the desired value.
Pmax
0.0 ... 30.0 MPa
Set maximum pressure limit for the
high pressure pump to the desired
value.
Pmin
0.0 ... 30.0
MPa
Set minimum pressure limit for the
high pressure pump to the desired
value.
Links
Die Registerkarte
Links
im Fenster
818 IC Pump
dient zur Wahl und
Einstellung der COM-Schnittstelle (Details siehe section 5.2.4 Links of
the «IC Net 2.3» Instructions for Use 8.110.8283).