3 Operation
818 IC Pump / 8.818.1023 Instructions for Use
20
switched off. At the same time, the
Pressure
field color in the 818 settings window is
changed to the value set for
Out of range
.
Send current parameters immediately to the
818 IC Pump (only available if pump is
running). Parameters are not stored in the
system file (
*.smt
) as long as the file is not
saved.
Set current parameters and start 818 IC Pump
drive.
Stop 818 IC Pump drive.
Time program
The
Program
tab of the
818 IC Pump
window a user-defined time program
can be entered. This program is started automatically as defined in the
Start mode
window (see section 4.4.3 «IC Net 2.3» Instructions for Use
8.110.8283), either at the moment the determination is started (
Start with
determination
) or at the moment the sample is injected (
Start with inject
).A
First column
Time at which program instruction is applied.
Entry range:
0.0 ... 999.9 min
If no time is entered, the program instruction is
applied together with the last instruction with
time entry.
Second column
Program instruction (see below).
Third column
Parameter for program instruction (see below).