Figure 22, Deaerating the high pressure pump – Metrohm 872 Extension Module IC Pump User Manual
Page 41
![background image](/manuals/586231/41/background.png)
■■■■■■■■■■■■■■■■■■■■■■
4 Installation
872 Extension Module IC Pump
■■■■■■■■
33
Deaerating the high pressure pump
The instrument must be connected to the PC and switched on to deaerate
the high pressure pump.
5
1
2
6
7
5
3
4
5
Figure 22
Deaerating the high pressure pump
1
Syringe 10 mL 6.2816.020
For aspirating the eluent.
2
Luer connector
On purging needle.
3
Purging needle 6.2816.040
4
Deaerating capillary
5
PEEK pressure screws, short
6.2744.070
6
Purge valve
7
Purge valve rotary knob
1 Connecting the purging needle
■
Push the end of the purging needle (22-3) over the end of the
deaerating capillary (22-4) on the purge valve.
2 Connecting the syringe
■
Insert syringe (22-1) in the Luer connector (22-2) of the purging
needle (see Figure 22, page 33).
3 Opening purge valve
■
Open the rotary knob (22-7) by approx.
½ rotation counterclock-
wise.