background image

■■■■■■■■■■■■■■■■■■■■■■

4 Installation

872 Extension Module IC Pump

■■■■■■■■

25

Assembling eluent aspiration tubing

1 Guide the free end of the eluent aspiration tubing (14-1) out of the

instrument through a suitable capillary feed-through.

2 Installing the 6.1602.160 eluent bottle attachment

Slide tubing nipple (14-2) and O-ring (14-3) onto the eluent aspi-
ration tubing (14-1).

Push eluent aspiration tubing (14-1) through the bottle attach-
ment (14-4) and screw tight.

1

2

3

4

Figure 14

Installing eluent bottle attachment

1

Eluent aspiration tubing 6.1834.080

2

Tubing nipple
From accessory set 6.1602.160.

3

O-ring
From accessory set 6.1602.160.

4

Bottle attachment
From accessory set 6.1602.160.

3 Mounting aspiration filter

Insert filter holder (15-1) into the aspiration filter (15-2) and screw
tight.

1

2

Figure 15

Mounting aspiration filter

1

Filter holder
From accessory set 6.2744.210.

2

Aspiration filter 6.2821.090