Figure 20, Install pump tubing connection with filter – Metrohm 872 Extension Module Liquid handling User Manual
Page 39

■■■■■■■■■■■■■■■■■■■■■■
5 Installation
872 Extension Module – Liquid handling
■■■■■■■■
31
3 Connecting the pressure side
NOTE
Depending on the use of the peristaltic pump, on the pressure side
you can either connect:
■
Case A: a 6.2744.180 pump tubing connection with filter
(see Figure 20, page 31) or
■
Case B: a 6.2744.160 pump tubing connection without filter
(see Figure 21, page 32).
For pumping the auxiliary solutions to the MSM or to the SPM, a
6.2744.180 pump tubing connection with filter must be used.
Case A: 6.2744.180 pump tubing connection with filter:
1
2
3
Figure 20
Install pump tubing connection with filter
1
Union nut
2
Adapter
3
Tubing olive with filter holder
■
Slide union nut (20-1) onto the pump tubing.
■
Select a suitable adapter (20-2) and slide it onto the pump tubing.
The type of adapter depends on the pump tubing (see Table 1,
page 32).
■
Place the tubing olive with filter holder (20-3) onto the pump tub-
ing.
■
Screw the union nut (20-1) onto the tubing olive (20-3).
or
Case B: 6.2744.160 pump tubing connection without filter: