beautypg.com

Introduction, Kit contents and specifications – Bird Technologies 7000A845 User Manual

Page 3

background image

1

Introduction

Connecting a mismatch of a known value and measuring it with the Antenna Tes-
ter provides a test to determine the degree of measurement accuracy.

Measurement accuracy is specified to be ± 0.040 of Rho (p), where:

Because of the manufacturing tolerance, each mismatch is analyzed across the
frequency band of the Antenna Tester. Rho data is recorded so that future com-
parisons can be made without regard to the manufacturing tolerance.

This verification kit is used to verify Antenna Tester performance. It is intended
for use with the AT-800 Antenna Tester. These instructions were written with the
understanding that the operator is familiarwith the operation of the Antenna Tes-
ter. Refer to the Antenna Tester Operator Manualfor more information.

Kit Contents and Specifications

Standard Mismatch—p/n 2-T-MN-1R50 (includes verification data)

VSWR

1.5:1 (nominal)

Rho ()

0.200 (nominal)

Connector

Type N (M)

Power Rating

1 Watt cw maximum

Operating Temperature

20° to 30°C (68° to 86°F)

Storage Temperature

0° to 50°C (32° to 122°F)

Adapter—p/n4240-403

Connector

N (F) to Universal Adapter (M)

Impedance

50

Ώ

nominal

Verification Data—Rho data included with mismatch,

Data Uncertainty

VSWR

± 2.4%

Rho

± 0.0121

ρ

VSWR 1

VSWR 1

+

-------------------------

=