beautypg.com

Figure 11 range adjustment potentiometers, Storage – Bird Technologies 4304A User Manual

Page 34

background image

23

20. Multiply by 10 the actual RF source power recorded in Step 15 (the reading

at “40” taken at 405 MHz). Set the RF source to this calculated value and the
frequency to 405 MHz.

21. Adjust R6 until the 4304A meter indicates 400.

Figure 11 Range Adjustment Potentiometers

Storage

When storing the Bird 4304A, turn the element so that the arrow points down
and set the range switch to 5 W. This will shunt the meter circuit and protect the
meter by dampening needle action.

3

9

8

7

2

4

1

10

+

R6

500

R7

150

R8

50

R9

15

R10

5